Detect Port Assignment
Starts a procedure by which the R&S
ZNB/ZNBT (with a little help from the attached
calibration unit) auto-detects the connected ports. The automatic assignment replaces
the configured one.
In case auto-detection fails
●
an error report is shown as a warning dialog
●
the undetected port connections are overlayed by warning signs
●
the calibration may be invalid
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select
"Default Port Assignment".
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Apply
Is enabled as soon as sufficient data have been acquired for the calibrated ports and
standards and for the power meter. The button starts the calculation of the calibration
data and closes the calibration wizard.
The results are applied to the active channel (or to all channels in the active recall set,
if all channels are calibrated).The current instrument settings are stored with the cor-
rection data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the
Use the
to perform an additional source flatness calibration (see
Chapter 4.5.7.3, "Combining SMARTerCal with Scalar Power Calibration"
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings may cause the calibration to be inaccurate. The analyzer auto-detects poten-
tial sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer