GUI Reference
R&S
®
ZNB/ZNBT
275
User Manual 1173.9163.02 ─ 38
2.
"Connectors and Cal Kits":
Select the connector type, gender and cal kit for all
ports to be calibrated. If necessary, import a calibration kit.
3.
"Calibration":
Acquire measurement data for the required ports or port pairs and
the required standards. Finally, decide whether or not to apply the resulting system
error correction.
Background and related information
Refer to
●
on page 139 for background information
●
Chapter 5.4.1.5, "Calibration Presetting Wizard (MultiCal)"
"multiple calibration" variant of the wizard
●
If the active channel is already equipped with a system error correction, the "Cali-
bration Presetting" wizard loads the underlying setup. If the calibration setup is not
changed and sweep data are available from previous calibrations, the existing sys-
tem error correction can be optimized without repeating the measurement of all
standards.
●
When you apply the acquired system error correction, the active calibration is
replaced and discarded.
To persist any kind of calibration you can transfer it to the "Cal Pool" using the
●
The active system error correction data can be read (and modified) using the
remote control command
Step 1: Ports and Type
Allows to select the test ports to be calibrated and the calibration type to be used.
Background information
Refer to
Chapter 4.5.1, "Calibration Types"
The layout of the main panel depends on the number of test ports available:
Cal Softtool