GUI Reference
R&S
®
ZNB/ZNBT
283
User Manual 1173.9163.02 ─ 38
Restart Sweep on Std. Meas.
If this function is active, a new standard measurement initiates a new sweep, starting
at the beginning ("Start") of the sweep range: The sweep points for the calibration
sweep are in ascending order, like for an ordinary measurement.
If "Restart Sweep on Std. Meas." is inactive, the new standard measurement is started
at the current sweep point; the current sweep is continued as a calibration sweep.
Note:
This function has a secondary effect in IDLE mode (i.e. while the calibration sweep is
NOT running):
●
If active, only the selected trace is refreshed.
●
If inactive, the R&S
ZNB/ZNBT permanently refreshes all traces of all diagrams
which may put a heavy load on the connected switch matrices (if any).
Hence it is recommended (and default) to activate it, in particular if one of the matrices
uses mechanical switches that will be worn off fast (apart from making noise).
Apply
Is enabled as soon as sufficient data have been acquired for the calibrated ports and
standards. The button starts the calculation of the system error correction and closes
the calibration wizard. The current instrument settings is stored with the correction
data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the "Calibration Manager" (see
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings may cause the calibration to be inaccurate. The analyzer auto-detects poten-
tial sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
[SENSe<Ch>:]CORRection:COLLect:DELete
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>:PORT<PhyPt>?
[SENSe<Ch>:]CORRection:DATE?
[SENSe<Chn>:]CORRection:SSTate?
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
5.4.1.5
Calibration Presetting Wizard (MultiCal)
setup and simultaneous manual acquisition of multiple system error corrections.
Access:
CHANNEL – CAL > "Start Cal" > "Start... (Manual)"
Cal Softtool