Setting parameters:
<Format>
LINear | LOGarithmic | DB | SIC
Keyword for the physical unit of the response values.
Default unit: dB (irrespective of the parameter selected)
Usage:
Setting only
CALCulate<Chn>:LIMit:LOWer:STATe
<Boolean>
CALCulate<Chn>:LIMit:UPPer:STATe
<Boolean>
These commands switch the lower and upper limit check on or off. Lower limit line seg-
ments are assigned even numbers; upper limit line segments are assigned odd num-
bers; see
CALCulate<Chn>:LIMit:LOWer[:DATA]
.
CALCulate<Chn>:LIMit:LOWer:STATe
does not affect
segments with odd numbers;
CALCulate<Chn>:LIMit:UPPer:STATe
does not
affect segments with even numbers.
Note:
to switch on or off the entire limit check,
including upper and lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - switch limit check on or off.
*RST:
OFF
Example:
CALC:LIM:LOW -10, 0, 0, -10
Define two limit line segments covering the entire sweep range.
Two upper limit line segments with default response values are
created in addition.
CALC:LIM:UPP 0, 5, 5, 0
Change the response values of the upper limit line segments.
CALC:LIM:LOW:STAT ON; :CALC:LIM:UPP:STAT ON; :
CALC:LIM:FAIL?
Switch the limit check on and query the result.
Manual operation:
See
CALCulate<Ch>:PARameter:DEFine
<TraceName>, <Result>[, <TestPortNum>]
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace is not displayed. To display a trace defined via
CALCulate<Ch>:PARameter:DEFine
, a window must be created (
ON
) and the trace must be assigned to this window
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED
); see example below.
Traces must be selected to become active traces; see
.
R&S ZVR/ZVAB Compatible Commands