GUI Reference
R&S
®
ZNB/ZNBT
516
User Manual 1173.9163.02 ─ 38
The limit line colors are defined in the "Define User Color Scheme" dialog ("SYSTEM >
DISPLAY > Config > Define User Color..."). You can choose between various options:
●
Display upper and lower limit lines with different colors.
●
Assign the same color to traces and associated limit lines.
●
Assign different colors to limit line segments with disabled limit check.
Note:
Display of the limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
CALCulate<Chn>:LIMit:DISPlay[:STATe]
Limit Check
Switches the limit check of the active trace on or off.
When the limit check is switched on, a movable "PASS" or "FAIL" message is dis-
played in the diagram. If the limit check fails at a measurement point, the point is
marked with a colored square. The "Limit Fail Trace" color is defined in the "Define
User Color Scheme" dialog ("Display > Display Config. > Define User Color..."). An
acoustic signal ("Global Beep") and a TTL signal indicating pass or fail can be gener-
ated in addition.
The appearance of the limit fail symbols is defined in the "Define User Color Scheme"
dialog ("SYSTEM > DISPLAY > Config > Define User Color..."). You can choose
between various options:
●
Change the trace color between failed measurement points.
●
Show or hide the colored squares.
Note:
Lines Softtool