Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
128
User Manual 1179.3315.02 ─ 01
Manual operation:
See
6.6.5
Configuring Sync/Burst
........................................................................................... 128
........................................................................................... 128
[SENSe:]DETect:REFerence
<Reference>
Sets the reference level to be used for setting the burst detection threshold.
Parameters:
<Reference>
ABSolute | NOISe | PEAK | RLEVel
*RST:
NOISe
Example:
SENSe:DETect:REFerence PEAK
Manual operation:
See
[SENSe:]DETect:THReshold
<Level>
Sets the detection threshold in dB/dBm relative to the burst detection reference level.
Parameters:
<Level>
<numeric value>
Default unit: dB
Example:
SENSe:DETect:THReshold -10
Manual operation:
See
6.6.6
Configuring Demodulation
..............................................................................................129
...................................................................................... 129
........................................................................................129
.........................................................................................129
.................................................................................... 129
[SENSe:]DEMod:MODE
<Mode>
Defines the demodulation mode (the demodulation standard).
Parameters:
<Mode>
HRP | BPRF | HPRF
*RST:
HRP
Example:
SENSe:DEMod:MODE HRP
Manual operation:
See
Configuring the Measurement