Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
120
User Manual 1179.3315.02 ─ 01
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
<TriggerLevel>
This command defines the level the external signal must exceed to cause a trigger
event.
Suffix:
<port>
.
Selects the trigger port.
1 = trigger port 1 (TRIGGER INPUT connector on front panel)
2 = trigger port 2 (TRIGGER INPUT/OUTPUT connector on front
panel)
3 = trigger port 3 (TRIGGER3 INPUT/OUTPUT connector on
rear panel)
Parameters:
<TriggerLevel>
Range:
0.5 V to 3.5 V
*RST:
1.4 V
Default unit: V
Example:
TRIG:LEV 2V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Configuring the Measurement