Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
117
User Manual 1179.3315.02 ─ 01
INPut<ip>:EATT
<Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (
INP:EATT:AUTO OFF
, see
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Attenuation>
attenuation in dB
Range:
see data sheet
Increment: 1 dB
*RST:
0 dB (OFF)
Default unit: DB
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation:
See
" Using Electronic Attenuation "
INPut<ip>:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
INP:EATT:AUTO OFF
Manual operation:
See
" Using Electronic Attenuation "
INPut<ip>:EATT:STATe
<State>
This command turns the electronic attenuator on and off.