Instrument tour
R&S
®
TSMExxDC
17
Manual 4901.1040.02 ─ 04
4 =
5 =
6 =
IF 1-5 (5x)
The IF 1 connector (blue frame) is the only IF output, which supports measure-
ments below 6 GHz. The IF1 is always connected to the master R&S TSMx6.
The connectors IF 2 to IF 5 are connected to additional devices. The synchroniza-
tion of the master device and the additional devices is done via the AUX connec-
tor.
Chapter 3.4, "Connecting to IF 1-5"
Pwr./State LEDs
Table 4-1: Pwr. and State LED states and their meaning
STATE LED
POWER LED
Meaning
off
off
no power supply connected at DC IN
power supply off
power supply < 10
V
off
yellow
standby
off
green, blinking
(2
Hz)
FGPA configuration in progress
off
green
R&S
TSMExxDC ready, not connected
green, blinking
rapidly
green
connected or measuring
green, blinking
2
Hz
green
Instrument is identified by the software
red, blinking 2
Hz green
temperature warning (controller board temperature =
65°
C to 70°
C)
red (continuous)
green
temperature error (controller board temperature above
70°
C)
*The fans are temperature-controlled and below a temperature of 48°
C on the
controller board, the fans are in status OFF.
AUX connector
The AUX connector (blue frame) is used to connect the master R&S TSMx6.
Rear panel view