Remote control commands
R&S
®
SMM100A
953
User Manual 1179.1341.02 ─ 05
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:POINts
<Points>
Sets the number of steps within the RF level sweep range.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Parameters:
<Points>
integer
Range:
2 to Max
Example:
See
Example "Setup an RF frequency or power sweep"
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the level sweep spacing. The sweep spacing for level sweeps is always linear.
Return values:
<Mode>
LINear
*RST:
LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result:
"LIN"
linear spacing
Usage:
Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
<Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Parameters:
<Logarithmic>
float
The unit
dB
is mandatory.
Range:
0.01 to 139 dB
Increment: 0.01
*RST:
1
Default unit: dB
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
SOURce subsystem