Remote control commands
R&S
®
SMM100A
941
User Manual 1179.1341.02 ─ 05
See
"RF frequency and level display with a downstream instrument"
Parameters:
<Power>
float
Level at the RF output, without level offset
Range:
See data sheet
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:POWer 15
Sets the level at RF output
Manual operation:
See
See
"Used SMM100A Settings For Measurement"
[:SOURce<hw>]:POWer:STARt
<Start>
[:SOURce<hw>]:POWer:STOP
<Stop>
Sets the RF start/stop level in sweep mode.
Parameters:
<Stop>
float
Sets the setting range calculated as follows:
(Lev OFFSet) to (Lev OFFSet)
Where the values are set with the commands:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
[:SOURce<hw>]:POWer:STARt
[:SOURce<hw>]:POWer:STOP
Range:
Minimum level to maximum level
*RST:
-30 (Start)/ -10 (Stop)
Default unit: dBm
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:POWer:STEP:MODE
<Mode>
Defines the type of step width to vary the RF output power step-by-step with the com-
mands
.
Parameters:
<Mode>
DECimal | USER
DECimal
Increases or decreases the level in steps of ten.
USER
Increases or decreases the level in increments, determined with
the command
[:SOURce<hw>]:POWer:STEP[:INCRement]
.
*RST:
DECimal
SOURce subsystem