Remote control commands
R&S
®
SMM100A
937
User Manual 1179.1341.02 ─ 05
Example:
POW:ALC:SLEV MIN
Sets the maximum attenuation.
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation
<Attenuation>
Sets the attenuation value of the RF signal in manual mode, set with command
.
Parameters:
<Attenuation>
integer
Range:
depends on the installed options
Increment: 4|5|6
*RST:
0
Example:
SOURce1:POWer:ATTenuation 20dB
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation:DIGital
<AttDigital>
Sets a relative attenuation value for the baseband signal.
Parameters:
<AttDigital>
float
Range:
-3.522 to 80
Increment: 1E-3
*RST:
0
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
<Mode>
Selects the state the attenuator is to assume if the RF signal is switched off.
Parameters:
<Mode>
UNCHanged | FATTenuation
FATTenuation
The step attenuator switches to maximum attenuation when RF
is off.
Note:
This setting overrides the RF level modes
.
UNCHanged
Freezes the current setting to keep the output impedance con-
stant during RF off.
*RST:
n.a. (preset & factory preset depends on option)
RFmax
≤
12.75
GHz: FATTenuation,
RFmax
>
12.75
GHz: UNChanged
SOURce subsystem