Remote control commands
R&S
®
SMM100A
697
User Manual 1179.1341.02 ─ 05
:OUTPut<hw>:AMODe
<AMode>
Sets the step attenuator mode at the RF output.
Parameters:
<AMode>
FIXed | MANual | AUTO
AUTO
The step attenuator adjusts the level settings automatically,
within the full variation range.
FIXed
The step attenuator and amplifier stages are fixed at the current
position, providing level settings with constant output VSWR.
The resulting variation range is calculated according to the posi-
tion.
*RST:
AUTO
Example:
:SOURce1:POWer:ALC:STATe 1
:OUTPut1:AMODe FIXed
Manual operation:
See
:OUTPut<hw>:BLANk:LIST:STATe
<State>
Activates RF output blanking.
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
1
Example:
:OUTPut1:BLAN:LIST:STAT ON
Activates the RF output blanking.
:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs.
Return values:
<Impedance>
G1K | G50 | G10K
*RST:
G50
Example:
:OUTPut1:IMP?
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage:
Query only
Manual operation:
See
OUTPut subsystem