Remote control commands
R&S
®
SMM100A
889
User Manual 1179.1341.02 ─ 05
[:SOURce<hw>]:FREQuency:MODE
<Mode>
Sets the frequency mode for generating the RF output signal. The selected mode
determines the parameters to be used for further frequency settings.
Parameters:
<Mode>
CW | FIXed | SWEep | LIST
CW|FIXed
Sets the fixed frequency mode.
CW
and
FIXed
are synonyms.
The instrument operates at a defined frequency, set with com-
mand
[:SOURce<hw>]:FREQuency[:CW|FIXed]
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:CENTer
[:SOURce<hw>]:FREQuency:MANual
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the
ter 13.16.11, "SOURce:LIST subsystem"
*RST:
CW
Example:
See
Example "Frequency configuration"
Example "Setup an RF frequency or power sweep"
Manual operation:
See
See
"State (RF frequency / level sweep)"
See
[:SOURce<hw>]:FREQuency[:CW|FIXed]
<Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
●
), the instrument operates at a fixed fre-
quency.
●
In sweep mode (
), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
SOURce subsystem