Remote control commands
R&S
®
SMM100A
730
User Manual 1179.1341.02 ─ 05
:SCONfiguration:EXTernal:REMote:CLEan
Removes all instruments from the pool of external instruments.
Example:
See
Example "Connecting and configuring external instruments"
Usage:
Event
Manual operation:
See
:SCONfiguration:EXTernal:CODer<ch>:REMote:SEND
<SendScpiCommand>
:SCONfiguration:EXTernal:FADer<ch>:REMote:SEND
<SendScpiCommand>
:SCONfiguration:EXTernal:RF<ch>:REMote:SEND
<SendScpiCommand>
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:SEND
<SendScpiCommand>
Allows you to send SCPI commands to the RF instruments connected to the
R&S
SMM100A.
Setting parameters:
<SendScpiCommand>"<SCPI syntax>"
String containing an SCPI command (query or setting)
Example:
See
Chapter 13.15, "SCONfiguration subsystem"
Usage:
Setting only
Manual operation:
See
:SCONfiguration:EXTernal:IQOutput<ch>:CONNections:CHECk?
Queries the status of the required connections between the R&S
SMM100A and the
R&S SZU.
R&S SZU is connected to the R&S
SMM100A via USB.
Example:
SCONfiguration:EXTernal:IQOutput1:REMote:ISELect "SZU (999991) B1066"
SOURce1:IQ:OUTPut:ANALog:TYPE SING
SCONfiguration:EXTernal:IQOutput1:REMote:CONNect
SCONfiguration:EXTernal:ACONnect 1
SCONfiguration:EXTernal:PBEHaviour 1
SCONfiguration:EXTernal:IQOutput1:REMote:INFO?
// "SZU (999991) B1066","USB (vendor specific)","999991","A"
SOURce1:FREQuency:CONVerter:EXTernal:CHECk:CONNections?
// Passed,Failed,Unused,Failed,Unused,Passed
// The R&S SZU is connected to the R&S SMW
// The USB and the RF A to LO In connections are correct
// There is a failure in the single-ended I/Q connection
// (between the I/Q OUT and the I/Q IN connectors)
Usage:
Query only
SCONfiguration subsystem