Network operation and remote control
R&S
®
SMM100A
628
User Manual 1179.1341.02 ─ 05
Chapter 14.5, "Resolving network connection failures"
For further information on the functions to read and write to an open session, and
the utility applications the software provides, see the R&S
VISA user manual.
Starting a remote control session over LAN with R&S VISA
To set the instrument to remote control, you can use the addressed command
>R
or
send any command from the controller.
1. Start the R&S
VISA Tester.
Establish the connection to the R&S
SMM100A.
"Setting up the controller with R&S VISA"
2. In the R&S VISA "Basics" tab, enter a SCPI command, e.g. "*IDN?". Confirm with
"Query".
The instrument is switched to remote control when it receives a command from the
controller.
3. Select "Read" to obtain the instrument response.
Tip:
If the "Show Log" checkbox is checked, R&S
VISA displays each VISA func-
tion call in the log-view on the left. If you check the "Write Log" checkbox, the log-
view entry is written to the log file as well. You can operate the log-view in two
modes: the "Live Mode" shows only the most recent messages whereas the "View
Mode" allows you to scroll the history.
4. To set, e.g. the frequency, enter
SOUR1:FREQ 4 GHz
and select "Write".
To check the performed setting,
SOUR1:FREQ?
and select "Query".
The instrument response is
4000000000
. The value corresponds to the frequency
in Hz.
Controlling the R&S
SMM100A remotely