Network operation and remote control
R&S
®
SMM100A
625
User Manual 1179.1341.02 ─ 05
used to replace the complete resource string. The host address is the
R&S
SMM100A's hostname or its IP address.
See also
Chapter 12.2.1, "LAN interface"
Setting up the controller with R&S VISA
To remote control the R&S
SMM100A, we use the R&S
VISA Tester application. The
application communicates via TCP/IP protocol.
The instrument is preconfigured for networks using DHCP (dynamic host configuration
protocol). If this configuration is used, enter the computer name in the position of the IP
address.
1. On the instrument:
a) Enable the LAN interface.
See
Chapter 12.7.1, "How to enable access via LAN"
b) Enable "SCPI over LAN".
See
Chapter 12.7.2, "How to activate LAN services"
2. On the controller (remote PC), install the R&S
VISA program.
http://www.rohde-schwarz.com/rsvisa
> "RS VISA Release Notes".
3. Connect the controller and the instrument in the same network (network cable).
Switch them on.
See also
Chapter 12.7.3, "How to connect to LAN"
4. On the controller, start "R&S
VISA > Tester 32bit" or "R&S
VISA > Tester 64bit".
Controlling the R&S
SMM100A remotely