a) Carefully align the connector of the cable and the connector along a common
axis.
b) Mate the connectors along the common axis until the male pin of the inner con-
nector engages with the female socket of the outer connector.
c) Turn the nut of the outer connector until the connectors are firmly coupled.
d) Torque the nut to the specified limit using a calibrated torque wrench. Hold the
opposite connector part stationary with a spanner.
For torque limits of the most relevant connector types, see
.
For more information, see chapter "Handling" of the application note
If your instrument is equipped with a test port adapter, see the application note
The connector types listed in this table represent the common connectors provided by
Rohde
&
Schwarz. It is considered as general information and therefore can contain
connector types that do not apply to your instrument.
See
Table 3-1: Connector types and torque limits
Type
Torque limit
Nut opening
lb-Inch
Nm
Inch
mm
N
13.3
1.5
3/4
20
SMA
5
0.56
5/16
8
3.5 mm
8
0.9
5/16
8
2.92 mm
8
0.9
5/16
8
2.4 mm
8
0.9
5/16
8
1.85 mm
8
0.9
5/16
8
1.0 mm
3
0.34
0.236
6
To prevent RF output switch-off
►
NOTICE!
If you set a too high output level without a load connected to the instru-
ment, the reverse power can exceed a limit forcing the R&S
SMM100A to switch off
the RF output.
Connect a load with sufficient return loss as given in the data sheet.
3.1.10
Connecting to LO In/Out
"LO In/Out" connectors are Subminiature Version A (SMA) connectors.
The connector is located on the
.
Follow the instructions in
"To connect to screwable connectors"
See also
Chapter 7.8, "Local oscillator coupling"
Preparing for use