How to use the internal adjustments
Deciding whether to run internal adjustments
1. Select "System Config > Setup > Internal Adjustment".
2. Observe the status and color indication in the section "Since Last Full Adjustment".
Green: internal adjustments are not required.
Red: internal adjustments are required. Observe also the indication in the "Informa-
tion" field.
3. We recommend that you run internal adjustments in the following cases:
● Before starting any application, that requires a maximum of level accuracy
● When a long period of time has passed since the last adjustments
● If the ambient temperature of the instrument significantly differs from the one of
the last adjustments
4. Proceed as described in
"Running internal adjustments"
Running internal adjustments
1.
NOTICE!
Adjustments can be invalid if performed when the instrument is not
warmed-up.
Wait until the instrument has reached its operating temperature before you start the
adjustment procedure.
The warm-up time is up to 30 minutes.
2. Select "System Config > Setup > Internal Adjustment > Adjust All".
The adjustment process starts.
If the internal adjustment has started, do not interrupt the process.
A progress indicator shows the status of the adjustment process. The extent of the
adjustments depends on the installed options.
If any error occurs, the process aborts, and an error message appears in the "Info"
line.
If there is an error:
● Observe the error message.
● Continue the process.
See
"Continuing the adjustment process if error appears"
Continuing the adjustment process if error appears
Per default, if any error occurs during the adjustment process, the process aborts. An
error message appears in the "Info" line.
If you prefer to finish the adjustments also if there is an error, proceed as follows:
1. Select "System Config > Setup > Security > Protection" an unlock the protection
level 1.
See
Performing maintenance tasks