Remote Control Commands
R&S
®
SMF100A
480
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:LFOutput:FREQuency:MANual
<Manual>
Determines the frequency and triggers the next sweep step manually in
, and
.
Note:
You can select any frequency within the setting range. The range is defined with
and
Parameters:
<Manual>
float
Range:
full frequency range
Increment: see the data sheet: Modulation sources > Internal
modulation generator > Resolution of frequency
setting
*RST:
1000
Example:
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step"
sweep mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:MODE
<Mode>
Sets the instrument operating mode, and determines the commands to be used for fre-
quency settings.
The source for the LF sweep is selected with command
.
SOURce Subsystem