Remote Control Commands
R&S
®
SMF100A
476
Operating Manual 1167.2319.02 ─ 12
7.13.7 SOURce:INPut Subsystem
The
SOURce:INPut
subsystem contains the commands for configuring the inputs for
external modulation signals. The instrument trigger setting influences all sweeps and is
effective in the List mode (Instrument Trigger).
[:SOURce<hw>]:INPut:MODext:COUPling<ch>
.................................................................476
[:SOURce<hw>]:INPut:MODext:IMPedance<ch>
...............................................................476
................................................................................... 477
[:SOURce<hw>]:INPut:MODext:COUPling<ch>
<Coupling>
The command selects the coupling mode for external feed via EXT 1 and/or EXT 2
input.
Parameters:
<Coupling>
AC | DC
AC
The DC voltage component is disconnected from the modulation
signal.
DC
The modulation signal is not changed.
*RST:
AC
Example:
SOUR:INP:MOD:COUP1 AC
selects the coupling mode AC for external modulation input
EXT1.
Options:
R&S SMF-B20
Manual operation:
See
" Coupling – External Inputs "
[:SOURce<hw>]:INPut:MODext:IMPedance<ch>
<Impedance>
Sets the impedance for external feed via the EXT 1 or EXT 2 input.
Parameters:
<Impedance>
G50 | G600 | G100k
G100k
100 kOhm to ground
G600
600 Ohm to ground
G50
50 Ohm to ground
*RST:
G100k
Example:
INP:MOD:IMP1 G50
the EXT input is set to 50 Ohm to ground.
Options:
R&S SMF-B20
Manual operation:
See
" Impedance – External Inputs "
SOURce Subsystem