Instrument Function
R&S
®
SMF100A
301
Operating Manual 1167.2319.02 ─ 12
Displays the pulse sequence as defined in the file.
"Edit"
Opens the pulse train dialog, see
. The dialog
graphically represents the pulse train signal and provides access to
the data editor.
"Zoom Position"
Sets the blue marker in the pulse train graph. The marker defines the
center of any zoom in or zoom out action.
"Zoom In / Zoom Out"
Enlarges the diagram by factor 2 per "Zoom In", or scales it down
accordingly when you select "Zoom Out".
Edit Pulse Train Data
Opens an editor allowing you to enter the "On-Time / OFF-Time" value pairs in a table.
In addition, you can assign a repetition rate to each pair. Based on these values, the
instrument then generates the pulse train signal. You can enter any number of value
pairs and save your list in a file. The file name is displayed in the header of the dialog.
"ON-Time/
μ
s"
Determines the length of the respective pulse (signal is high).
Remote command:
[:SOURce<hw>]:PULM:TRAin:ONTime
"OFF-Time/
μ
s" Determines the time length, the signal level of the pulse remains low.
Remote command:
[:SOURce<hw>]:PULM:TRAin:OFFTime
"Count"
Sets the number of repetitions for each pulse ("ON-Time/
μ
s"/"OFF-
Time/
μ
s" value pair).
Tip:
If you set "Count = 0", the corresponding value pair is ignored in
the pulse sequence. With this function you can skip value pairs indi-
vidually, without deleting them from the table. This allows re-enabling
a value pair by entering a number unequal to zero.
Remote command:
[:SOURce<hw>]:PULM:TRAin:REPetition
"Goto"
Selects row for editing.
If Goto row is selected, a window opens for entering the requested
row.
(it is not possible to change individual positions of the list)