Instrument Function
R&S
®
SMF100A
248
Operating Manual 1167.2319.02 ─ 12
"Insert Range"
Insert new rows before the marked row. The number of rows to be
inserted can be defined in an entry window.
"Fill...."
Opens a sub menu for defining a set of list values to be automatically
entered in the ucor list (see
Chapter 5.4.7.2, "Filling the Correction
"Fill With Sen-
sor"
Calls the menu to activate the filling of the user correction list with
level values acquired by the selected power sensor (see
ter 5.4.7.3, "Filling the Correction List with Power Sensor Measure-
ment Data"
"Delete Row"
Deletes the marked row.
"Delete
Range..."
Allows to delete any number of rows starting with the marked row.
The number of rows to be deleted can be defined in an entry window.
"Save as"
Open the file menu to save the list under a new name.
Note:
Save list only after filling both columns (frequency and level),
otherwise the entries are lost.
Each list is saved to the R&S
SMFCompactFlash™ Card as a sepa-
rate file with the file prefix
*.uco
. The file name and the directory to
which the file is saved are user-selectable.
"Save"
The list is saved under its current name.
Remote command:
[:SOURce<hw>]:CORRection:CSET[:SELect]
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
Import/Export
User correction list can be imported from externally created files or exported into text or
CSV-files. The import/export settings are available after clicking the "Import/Export"
button.
Import/Export >>>
Expands the menu with the area for import and export of user correction files.
Externally edited Excel tables with any number of frequency/level value pairs can be
imported as text or CSV-files and used for user correction.
Conversely, you can also export internally created user correction lists as text or CSV-
files.