Remote Control Commands
R&S
®
SMF100A
455
Operating Manual 1167.2319.02 ─ 12
Example:
CORR:CSET '/var/user/'
selects the table ucor1.
CORR:CSET:DATA:FREQ:POIN?
queries the number of frequency values in the table ucor1.
Response:
440
the table ucor1 contains 440 frequency values.
Usage:
Query only
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
<Power>
Transfers the level data to the table selected with
.
*RST
does not affect data lists. The numerical suffix at
SOURce
must not be used for
this command.
Parameters:
<Power>
Power#1[, Power#2, ...]
Example:
CORR:CSET '/var/user/ucor1'
selects the table
ucor1
.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
enters the level values in the table
ucor1
.
Manual operation:
See
" Edit User Cor. Data - User Correction "
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of level values in the selected table.
The numerical suffix at
SOURce
must not be used for this command.
Return values:
<Points>
integer
Range:
0 to 10000
*RST:
0
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW:POIN?
queries the number of level values in the table ucor1.
Response:
440
the table ucor1 contains 440 level values.
Usage:
Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
The command fills the selected user correction list with the level values measured by
the power sensor for the given frequencies.
To select the used power sensor set the suffix in key word
SENSe
.
SOURce Subsystem