Instrument Function
R&S
®
SMF100A
119
Operating Manual 1167.2319.02 ─ 12
The user-programmable marker signal can be used for brightness control of an oscillo-
scope. The output becomes active when the sweep run has reached the marker. Up to
10 markers can be set to mark positions in the sweep run. The duration of the active
signal is equal to the dwell time (DWELL) of a step.
Remote command:
[:SOURce<hw>]:SWEep:MARKer:OUTPut:POLarity
Blank Polarity
Selects the polarity of the Blank marker.
Remote command:
Blank Width
Selects the start of the blank marker in relation to the external sweep trigger (blank
width).
"RF Blanking"
The blank marker starts after the external sweep trigger signal.
"Trigger and
RF"
The blank marker starts with the external sweep trigger signal.
Remote command:
RF Blanking
Selection of the RF blanking method for sweeps with step widths less than 10 MHz.
"Auto"
RF blanking only when the step synthesizer switches to the next step
(provides a clean RF spectrum).
"On"
RF blanking after each step.
"Off"
No RF blanking.
Remote command:
Analyzer Type
Allows to switch the HP8757D/E scalar network analyzer (selection HP8757D/E).
Note:
The HP8757D/E has a special control mechanism via GPIB bus. RF blanking is
internally set to "Auto". Start and stop frequencies are transferred to the network ana-
lyzer. An GPIB command is not available, since it is used for controlling the network
analyzer.
General Instrument Settings