Remote Control Commands
R&S
®
SMC100A
288
Operating Manual 1411.4060.32 ─ 04
The commands of this subsystem are not affected by an instrument reset (
..................................................................288
...............................................................288
..............................................................288
............................................................289
.....................................................................................289
[:SOURce]:ROSCillator:EXTernal:FREQuency
<Frequency>
The command informs the instrument of the frequency of the external reference.
Parameters:
<Frequency>
5MHz | 10MHz
Default unit: Hz
Example:
ROSC:SOUR EXT
selects the external source. The reference must be input at the
REF IN input.
ROSC:EXT:FREQ 5MHz
informs the instrument that the external reference frequency is
5MHz.
Usage:
SCPI confirmed
[:SOURce]:ROSCillator:EXTernal:RFOFf:STATe
<State>
The command determines if the RF output is switched off in case of a missing external
reference signal for selection external source.
If enabled, this setting ensures that no improper RF signal due to the missing external
reference signal is output and used for measurements.
In addition to the error message "Ext Ref missing", message "RF output deactivated' is
generated.
Parameters:
<State>
0 | 1 | OFF | ON
Example:
ROSC:SOUR EXT
selects the external source. The reference must be input at the
REF IN input.
Example:
ROSC:EXT:RFOF:STAT ON
In case of a missing external signal, no RF signal is output.
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue
<Value>
The command specifies the frequency correction value (adjustment value).
SOURce Subsystem