Remote Control Commands
R&S
®
SMC100A
257
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Fixed>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
1GHz
Default unit: Hz
Example:
FREQ 500kHz
sets the frequency of RF output signal A to 500 kHz.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency[:CW|:FIXed]:RCL
<Rcl>
The command determines whether the current frequency setting is retained or whether
the stored frequency setting is adopted when an instrument configuration is loaded.
*RST does not affect this setting.
Parameters:
<Rcl>
INCLude | EXCLude
INCLude
The stored frequency is also loaded when a stored instrument
configuration is loaded.
EXCLude
The RF frequency is not loaded when a stored instrument config-
uration is loaded. The current frequency is retained.
*RST:
no default
Example:
FREQ:RCL INCL
The stored frequency is set if the Recall command is called.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MANual
<Manual>
In Sweep mode (
:SOUR:FREQ:MODE SWE
) the command sets the frequency for the next
sweep step in the "Step" sweep mode (
SOUR:SWE:MODE MAN
). Here only frequency
values between the settings
[SOUR]:FREQ:STAR
and
...:STOP
are permitted. Each
sweep step is triggered by a separate
SOUR:FREQ:MAN
command.
As with the "Frequency" value entered in the header, the
OFFSet
value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet
is set to 0. The value range for other
OFFset
values can be calculated using
the following formula:
STARt
+
OFFSet
...
STOP
+
OFFSet
SOURce Subsystem