Remote Control Commands
R&S
®
SMC100A
232
Operating Manual 1411.4060.32 ─ 04
7.10 OUTPut Subsystem
The
OUTPut
system contains the commands which set the properties of the RF output
connector.
The properties of the LF output connector are set in the
chapter 7.12.6, "SOURce:LFOut-
.............................................................................232
..............................................................................232
..............................................................................................233
...................................................................................233
................................................................................233
...........................................................................................234
:OUTPut<hw>:AFIXed:RANGe:LOWer?
The command queries the minimum level which can be set without the attenuator being
adjusted (Attenuator FIXed).
Return values:
<Lower>
float
Default unit: dBm
Example:
OUTP:AFIX:RANG:LOW
queries the minimum level for the FIXed setting.
Example:
Response: -50
The minimum level is -50 dBm.
Usage:
Query only
:OUTPut<hw>:AFIXed:RANGe:UPPer?
The command queries the maximum level which can be set without the attenuator being
adjusted (Attenuator FIXed).
Return values:
<Upper>
float
Default unit: dBm
Example:
OUTP:AFIX:RANG:UPP
queries the maximum level for the FIXed setting for the RF output.
Example:
Response: -27
The maximum level is -27 dBm.
Usage:
Query only
OUTPut Subsystem