Remote Control Commands
R&S
®
SMBV100B
913
User Manual 1178.4460.02 ─ 03
:SYSTem:RESTart
Restarts the instrument without restarting the operating system.
Usage:
Event
:SYSTem:SHUTdown
Shuts down the instrument.
Usage:
Event
Manual operation:
See
:SYSTem:WAIT
<TimeMs>
Delays the execution of the subsequent remote command by the specified time.
This function is useful, for example to execute an SCPI sequence automatically but
with a defined time delay between some commands.
See
Chapter 10.3.3, "Assigning Actions to the [
Setting parameters:
<TimeMs>
integer
Wait time in ms
Range:
0 to 10000
*RST:
0
Example:
:SYSTem:WAIT 10000
// waits 10s before resetting the instrument
*RST
Usage:
Setting only
:SYSTem:FILes:TEMPorary:DELete
Deletes the temporary files from the internal memory or, if installed, from the Non-Vola-
tile Memory.
Example:
:SYSTem:FILes:TEMPorary:DELete
// temporary files are removed
Usage:
Event
12.18
STATus Subsystem
This system contains the commands for the status reporting system. See also
ter A.1.5, "Status Reporting System"
on page 965 for detailed information.
on page 613 has no effect on the status registers.