![Rohde & Schwarz SMBV100B User Manual Download Page 883](http://html.mh-extra.com/html/rohde-and-schwarz/smbv100b/smbv100b_user-manual_1477881883.webp)
Remote Control Commands
R&S
®
SMBV100B
881
User Manual 1178.4460.02 ─ 03
Options:
DER1G requires R&S
SMBVB-B3
Manual operation:
See
"Reference Output/1GHz Reference Output"
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue
<Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value>
integer
*RST:
0
Example:
See
[:SOURce]:ROSCillator[:INTernal]:ADJust[:
Manual operation:
See
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
<State>
Determines whether the calibrated (off) or a user-defined (on)
adjustment value
is
used for fine adjustment of the frequency.
Parameters:
<State>
0 | 1 | OFF | ON
0
Fine adjustment with the calibrated frequency value
1
User-defined adjustment value.
The instrument is no longer in the calibrated state.
The calibration value is, however, not changed. The instrument
resumes the calibrated state if you send
SOURce:ROSCillator:INTernal:ADJust:STATe 0
.
*RST:
n.a. (factory preset: 0)
Example:
Example "Configuring the reference oscillator"
Manual operation:
See
12.16.17
SOURce:SWEep Subsystem
The
SOURce:SWEep
subsystem contains the commands for configuring RF sweep sig-
nals.
SOURce Subsystem