Remote Control Commands
R&S
®
SMBV100B
835
User Manual 1178.4460.02 ─ 03
Setting parameters:
<Filename>
"<filename>"
Filename or complete file path
Usage:
Setting only
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE
<Type>
Sets the type of the analog signal.
Parameters:
<Type>
SINGle | DIFFerential
*RST:
SING (if SCONfiguration:EXTernal:PBEHaviour 0) /
- (if SCONfiguration:EXTernal:PBEHaviour 1)
Example:
SOURce1:IQ:OUTPup:ANALog:TYPE DIFFerential
Options:
DIFFerential requires R&S
SMBVB-K17
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE
<Mode>
Determines the mode for setting the output parameters.
Parameters:
<Mode>
FIXed | VARiable | VATTenuated
FIXed
Locks the I/Q output settings
VARiable
Unlocks the settings
VATTenuated
Attenuates the signal with the value set with the command
SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:
DATTenuation
*RST:
FIXed
Example:
See
Example "Enabling and configuring the analog output"
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut:LEVel
<Level>
Sets the off-load voltage Vp of the analog I/Q signal output.
The value range is adjusted so that the maximum overall output voltage does not
exceed 4V, see
"Maximum overall output voltage"
SOURce Subsystem