Multi-Instrument Setups
R&S
®
SMBV100B
466
User Manual 1178.4460.02 ─ 03
5. To configure connected external instruments, use the "System Config > System
Configuration >
External RF and I/Q" dialog.
The parameters defining the voltage level of the output signal are interdependent.
Consider the limits listed in
"Maximum overall output voltage"
Maximum overall output voltage
The voltage of the analog output signals is defined as a combination of output voltage
of the I and Q signal components and an optional bias voltage. If the differential output
is used, an additional offset can be set, too. The values of these parameters are inter-
dependent so that the magnitude of the sum of output voltage and bias voltage must
not exceed 4V.
The following applies:
●
Single ended signal:
V
p
+ V
bias
≤
4
V
●
Differential signal:
0.5*V
p
+ |V
bias
| + 0.5*V
offset
≤
4
V
Where:
●
V
p
is the output voltage as set with the parameter
●
V
bias
is the bias voltage as set with the parameter
●
V
offset
is the offset between inverting and non-inverting output as set with the
For more information, see data sheet.
Related settings
●
Chapter 8.7.1, "General Analog I/Q Output Settings"
●
Chapter 4.2, "Impairing the Signal"
See user manual R&S
SMBVB-K540, R&S
SMBVB-K541 Envelope Tracking and
AM/AM, AM/PM Predistortion.
8.7.1
General Analog I/Q Output Settings
Access:
► Select "I/Q Analog > I/Q Analog Outputs > General".
For description of the impairments settings, see
Chapter 4.2, "Impairing the Signal"
Settings:
I/Q Analog Output Settings