Remote Control Commands
R&S
®
SMBV100B
705
User Manual 1178.4460.02 ─ 03
SOURce:PULM:STATe 1
OUTPut1:STATe 1
The following commands are available:
........................................................................................ 705
[:SOURce<hw>]:PULM:TRIGger:MODE
........................................................................... 705
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate
................................................... 706
....................................................................................... 706
........................................................................................ 706
........................................................................................ 706
[:SOURce<hw>]:PULM:DOUBle:DELay
............................................................................707
[:SOURce<hw>]:PULM:DOUBle:WIDTh
........................................................................... 707
[:SOURce<hw>]:PULM:DOUBle:STATe
............................................................................ 707
.........................................................................................707
..................................................................................... 707
........................................................................................ 708
[:SOURce<hw>]:PULM:OUTPut:VIDeo:POLarity
............................................................... 708
..................................................................................... 708
..................................................................................709
[:SOURce<hw>]:PULM:MODE
<Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode>
SINGle | DOUBle
SINGle
Generates a single pulse.
DOUBle
Generates two pulses within one pulse period.
*RST:
SINGle
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:TRIGger:MODE
<Mode>
Selects a trigger mode - auto, single, external, external single or external gated - for
generating the modulation signal.
Parameters:
<Mode>
AUTO | EXTernal | EGATe | SINGle | ESINgle
*RST:
AUTO
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
SOURce Subsystem