Example:
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
// 1
SOURce:FREQuency:CW?
// 60000000000
SOURce1:IQ:OUTPup:ANALog:TYPE DIFFerential
SCONfiguration:EXTernal:PBEHaviour
1
*RST
SOURce:FREQuency:CW?
//
60000000000
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
//
1
SOURce1:IQ:OUTPup:ANALog:TYPE?
//
DIFF
SCONfiguration:EXTernal:PBEHaviour
0
*RST
SOURce:FREQuency:CW?
//
1000000000
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
//
0
SOURce1:IQ:OUTPup:ANALog:TYPE?
//
SING
Manual operation:
See
"Preset behavior: Keep connections to external instru-
:SCONfiguration:EXTernal:DIGital:DIRection?
:SCONfiguration:EXTernal:IQOutput:DIRection?
Queries the connector direction.
Return values:
<Direction>
NONE | IN | OUT
*RST:
IN
Example:
See
Example "Connecting and configuring external instrument"
Usage:
Query only
Manual operation:
See
:SCONfiguration:EXTernal:DIGital:IQConnection:STATe?
Queries the status of the I/Q connection of the digital interfaces.
Return values:
<IQConnState>
0 | 1 | OFF | ON
Example:
See
Example "Connecting and configuring external instrument"
SCONfiguration Subsystem