Manual operation:
See
" Sensor type and serial number "
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note:
Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
SENS1:ZERO
Executes zeroing.
Usage:
Event
Manual operation:
See
12.15
SCONfiguration Subsystem
The
SCONfiguration
subsystem contains the commands for defining the system
configuration setting, like stream mapping and used baseband configuration.
Required options
See
Example: Distributing the I/Q streams to the output connectors
The following example lists the commands necessary to fulfill this task.
// ******************************************************************
// Map the I/Q streams to the output connectors
// ******************************************************************
SCONfiguration:OUTPut:MAPPing:RF1:STReam1:STATe ON
SCONfiguration:OUTPut:MAPPing:IQOutput1:STReam3:STATe ON
Example: Connecting and configuring external instrument
The following example lists the commands necessary to fulfill this task.
SCONfiguration:EXTernal:DISPlay ALL
SCONfiguration:EXTernal:REMote:CLEan
SCONfiguration:EXTernal:REMote:SCAN
SCONfiguration:EXTernal:REMote:SCAN:STATe?
// 0
SCONfiguration Subsystem