
Getting Started
R&S
®
SMBV100B
61
User Manual 1178.4460.02 ─ 03
signals and marker signals specific to the firmware option. With suitable marker set-
tings for instance, you can mark slot or frame boundaries or mark the start of a particu-
lar modulation symbol.
This example extends further the configurations performed in
ing a Digitally Modulated Signal"
on page 53. We assume a default connector map-
ping (see
).
This test setup requires one oscilloscope, like the R&S
®
RTO, as additional equipment.
1. In the block diagram, select "Block Diagram > Baseband > Misc > Custom Digital
Mod > Marker" tab.
2. Select "Marker Mode > Marker 1 > Pulse" and "Divider = 32".
Generated is a periodic marker with marker frequency of 120 KHz. The signal is
output at the User 1 connector of the R&S
3. Use a suitable cable to connect the User 1 connector of the R&S
SMBV100B to the
monitoring instrument, for example oscilloscope like R&S
®
RTO. See
Figure 2-10: Simplified representation of a test setup for signal monitoring**
** = The figure depicts the cabling as a general principle
4. Use a suitable cable to connect the
connector of the R&S
SMBV100B to the
monitoring instrument.
To learn more about this topic, refer to
Chapter 3.4.1.2, "Regular Marker Output Sig-
2.3.5
Verifying the Generated Signal with the Graphics Display
It is often useful to check the spectra of the configured signals, before you enable the
RF output of the instrument.
Trying Out the Instrument