Getting Started
R&S
®
SMBV100B
32
User Manual 1178.4460.02 ─ 03
Mounting in a rack
The R&S
SMBV100B can be installed in a rack using a rack adapter kit (Order No. see
data sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to
ensure that the instruments do not overheat. Insufficient airflow for a longer period can
disturb the operation and even cause damage.
2.1.1.5
Connecting AC Power
The R&S
SMBV100B is equipped with an AC power supply connector, that can be
operated with different AC power voltages. Once it is connected, the instrument auto-
matically adjusts to the given voltage. Refer to the data sheet for the requirements of
voltage and frequency. There is no need to set the voltage manually or change fuses.
To connect the AC supply
► Connect the R&S
SMBV100B to the AC power source using the supplied power
cable.
Note:
Since the instrument is designed in compliance with standard EN 61010-1
safety class I, it must only be connected to an outlet that has a ground contact.
Characteristics of the AC power supply:
●
100 V to 240 V AC
●
Max 3.5 A to 1.6 A (50 Hz to 60 Hz)
●
Max 3.5 A to 2.9 A (400 Hz)
2.1.1.6
Turning the Instrument On and Off
To turn on the R&S
SMBV100B
1.
Connect the instrument to the AC supply
2. Turn on the main AC power switch at the rear panel of the R&S
SMBV100B (posi-
tion "I" (on)).
The instrument is supplied with AC power.
Warm-up time for OCXO
When the instrument is switched on, the OCXO requires an extended warm-up time
(see data sheet).
Preparing for Use