Parameters:
<Shape>
SAWTooth | TRIangle
SAWTooth
One sweep runs from start to stop frequency. Each subsequent
sweep starts at the start frequency, i.e. the shape of the sweep
sequence resembles a sawtooth.
TRIangle
One sweep runs from start to stop frequency and back, i.e. the
shape of the sweep resembles a triangle. Each subsequent
sweep starts at the start frequency.
*RST:
SAWTooth
Example:
SOUR:SWE:SHAP TRI
selects the sweep cycle with alternating ascending and
descending sweep directions.
Manual operation:
See
" Shape - RF Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
<Spacing>
Selects the mode for the calculation of the frequency sweep intervals. The frequency
increases or decreases by this value at each step.
The keyword
[:FREQuency]
can be omitted. Then the command is SCPI-compliant.
Parameters:
<Spacing>
LINear | LOGarithmic
LINear
With the linear sweep, the step width is a fixed frequency value
which is added to the current frequency. The step width for lin-
ear sweep is entered in Hz (see
LOGarithmic
With the logarithmic sweep, the step width is a constant fraction
of the current frequency. This fraction is added to the current fre-
quency. The logarithmic step width is entered in % (see
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
on page 812).
*RST:
LINear
Example:
SWE:SPAC LIN
selects linear sweep spacing for a frequency sweep at the RF
output.
Manual operation:
See
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
<Linear>
Sets the step size for linear RF frequency sweep steps.