Example:
BB:MCCW:TRIG:OUTP2:PULS:DIV 4
sets the divider for the marker signal on output MARKER 2 to
the value 4.
BB:MCCW:TRIG:OUTP2:MODE PULS
enables the pulsed marker signal
BB:MCCW:TRIG:OUTP2:PULS:FREQ?
queries the pulse frequency for the marker signal.
Response: "33 000"
the resulting pulse frequency is 33 kHz.
Usage:
Query only
Manual operation:
See
Clock Settings
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE
...................................................................... 713
[:SOURce<hw>]:BB:MCCW:CLOCk:MULTiplier
................................................................ 713
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce
................................................................... 714
[:SOURce<hw>]:BB:MCCW:CLOCk:SYNChronization:MODE
............................................ 714
[:SOURce<hw>]:BB:MCCW:CLOCk:SYNChronization:EXECute
.........................................715
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE
<Mode>
Sets the type of externally supplied clock (
:BB:MCCW:CLOCk:SOURce EXTernal
).
When
MSAMple
is used, a multiple of the sample clock is supplied via the CLOCK con-
nector and the sample clock is derived internally from this. The multiplier is entered
with the command
BB:MCCW:CLOCk:MULTiplier
.
Parameters:
<Mode>
SAMPle | MSAMple
*RST:
SAMPle
Example:
BB:MCCW:CLOC:MODE SAMP
selects clock type "Sample", i.e. the supplied clock is a sample
clock.
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:CLOCk:MULTiplier
<Multiplier>
Specifies the multiplier for clock type Multiple Samples (
:BB:MCCW:CLOCk:MODE
MSAM
) in the case of an external clock source.
Parameters:
<Multiplier>
integer
Range:
1 to 64
*RST:
4