Remote Control Commands
R&S
®
SMBV100A
702
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt]
<Start>
Sets the power for the individual carriers in the defined carrier range. If the com-
mand
:BB:MCCW:EDIT:CARR:POW:STEP
is used to define a step width, the power
entered here applies only to the starting carrier. The power of the remaining carriers is
stepped up or down by the power specified in the
:BB:MCCW:EDIT:CARR:POW:STEP
command.
Parameters:
<Start>
float
Range:
-80 to 0
Increment: 0.01
*RST:
0
Example:
BB:MCCW:EDIT:CARR:POW -50 dB
sets the power of the carrier to -50 dB.
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STARt
<Start>
Selects the first carrier in the carrier range to which the settings with
the
:BB:MCCW:EDIT:CARR:...
commands shall apply.
Parameters:
<Start>
integer
Range:
0 to 8191
*RST:
0
Example:
BB:MCCW:EDIT:CARR:STAR 4
the carrier range starts at carrier 4.
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STATe
<State>
Switches all the carriers in the selected carrier range on or off.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
BB:MCCW:EDIT:CARR:STAT ON
sets all the carriers in the carrier range to ON.
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STOP
<Stop>
Selects the last carrier in the carrier range to which the settings with
the
:BB:MCCW:EDIT:CARR:...
commands shall apply.
SOURce Subsystem