Instrument Function
R&S
®
SMBV100A
441
Operating Manual 1176.8016.02 ─ 17
Carrier Stop
Defines the stop index of the carrier range to which the following settings are intended
to apply.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STOP
Carrier State
Switches the carriers in the carrier range on/off.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STATe
Power Start
Sets the power of the starting carrier.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt]
Power Step
Sets the width of the step with which the power will be changed from carrier to carrier.
The carrier power that is set with
Power + n* Power Step
must be within the valid
value range -80 dB to 0 dB.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer:STEP
Phase Start
Sets the phase of the starting carrier. This setting is only available for "Optimize Crest
Factor Mode" = Off.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe[:STARt]
Phase Step
Sets the width of the step with which the phase will be changed from carrier to carrier.
The phase that is set with
Phase + n* Phase Step
must be within the valid value range
0° to 360°.
This setting is only available for "Optimize Crest Factor Mode" = Off.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe:STEP
Accept
Adopts the carrier range setting into the table ("Carrier Table").
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:EXECute
Carrier Table and Carrier Graph
The lower part of the menu is used to open submenus for configuring carrier tables and
for setting triggers, markers and clocks.
Baseband Signal - Baseband Block