Instrument Function
R&S
®
SMBV100A
146
Operating Manual 1176.8016.02 ─ 17
Set to default
Enables you to reset the
*IDN
and
*OPT
strings in user defined mode, see
The default strings vary depending on the selected emulation mode (
Remote command:
IDN String
Indicates the identification string of the instrument when queried with the common
command
*IDN?
.
To assign a user defined identification string, select
Remote command:
on page 834 (user defined mode)
OPT String
Indicates the option string of the instrument as queried with the common command
*OPT?
.
In "User defined" (see
, you can create a user defined option string in
addition to the automatically created one.
Remote command:
5.2.3.17
Protection
This "Protection" dialog provides access to the unlocking of different protection levels .
Access:
► Select "Setup > Protection"
After power on the instrument, all protection levels are locked. To unlock the protec-
tion, the correct password must be entered, see
"To unlock or lock a protection level..."
The following functions are protected in the respective levels:
●
Protection Level 1
General Instrument Settings