Remote Control Commands
R&S
®
SMBV100A
717
Operating Manual 1407.6062.32 ─ 08
<password>
The respective functions are disabled when the protection level is
activated. No password is required for activation. A password must
be entered to deactivate the protection level. The password for the
first level is 123456. This protection level can be used to lock-out
internal adjustments.
Example:
SYST:PROT1 ON
activates protection level 1. Internal adjustments are only possible
after deactivating the lock-out.
SYST:PROT1 OFF, 123456
deactivates protection level 1. Internal adjustments are enabled
again.
Usage:
SCPI confirmed
SYSTem:SERRor?
This command returns a list of all errors existing at the time when the query is started.
This list corresponds to the display on the info page under manual control.
Return values:
<Serror>
string
Example:
SYST:SERR
queries all errors existing in the error queue.
Example:
Response: -221, 'Settings conflict', 153, 'Input
voltage out of range'
The two returned errors have occurred since the error queue was
last queried.
Usage:
Query only
:SYSTem:STARtup:COMPlete?
The command queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP
Response: 1
The startup of the instrument is completed.
Usage:
Query only
:SYSTem:TIME
<Time>
The command queries / sets the time for the instrument-internal clock.
The parameter is protected, in order to prevent accidental changes. It can be accessed
with protection level 1, see
SYSTem Subsystem