Remote Control Commands
R&S
®
SMBV100A
629
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Rcl>
INCLude | EXCLude
INCLude
The stored frequency is also loaded when a stored instrument
configuration is loaded.
EXCLude
The RF frequency is not loaded when a stored instrument config-
uration is loaded. The current frequency is retained.
*RST:
no default
Example:
FREQ:RCL INCL
The stored frequency is set if the Recall command is called.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MANual
<Manual>
In Sweep mode (
:SOUR:FREQ:MODE SWE
) the command sets the frequency for the next
sweep step in the "Step" sweep mode (
SOUR:SWE:MODE MAN
).
The parameter
<Manual>
must be a frequency value between the settings
[SOUR]:FREQ:STAR
and
...:STOP
. Each sweep step is triggered by a separate
SOUR:FREQ:MAN
command.
UP
or
DOWN
trigger the next sweep step in direction of higher
or lower frequencies.
As with the "Frequency" value entered in the header, the
OFFSet
value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet
is set to 0. The value range for other
OFFset
values can be calculated using
the following formula:
STARt
+
OFFSet
...
STOP
+
OFFSet
Parameters:
<Manual>
float
Range:
STARt to STOP
Increment: 0.01Hz
*RST:
100MHz
Default unit: Hz
Example:
SWE:MODE MAN
sets the
Step
sweep mode.
Example:
FREQ:MODE SWE
sets the Frequency Sweep mode. The sweep start frequency is
output.
FREQ:MAN UP
triggers the next higher sweep step.
FREQ:MAN 500MHz
outputs 500 MHz RF frequency (must e within the sweep fre-
quency range).
FREQ:MAN DOWN
triggers the next lower sweep step relative to 500 MHz.
SOURce Subsystem