MMEM:DCAT:LENG? '/usb/user'
Response:
2
2. To query the number of files in the current or specified directory, perform:
MMEM:CAT:LENG? '/usb/user'
Response:
3
3. Create a new subdirectory for mass memory storage in the specified directory.
MMEM:MDIR '/usb/user/new'
4. Copy the file
state
to a new file.
MMEM:COPY '/var/smbv/user/state.savrcltxt','/usb/user/new'
5. Rename the file
state
.
MMEM:MOVE 'state.savrcltxt','state_new.savrcltxt'
6. Remove the
test
directory.
MMEM:RDIR '/usb/user/test'
7.11.4 Remote Control Commands
...........................................................................................478
........................................................................................480
................................................................................................481
..............................................................................................482
:MMEMory:CATalog?
<path>
Returns the content of the current or a specified directory.
Query parameters:
<path>
string
String parameter to specify the directory. If the direcory is omitted,
the command queries the content of the current directory, queried
with
MMEM:CDIR
command.