Instrument Function
R&S
®
SMBV100A
163
Operating Manual 1407.6062.32 ─ 08
Mode - LO Coupling
Selects the mode of the local oscillator coupling.
"Internal"
This mode corresponds to a normal operation.
The internal local oscillator is used.
"External"
An external signal is used.
Note:
Selection of LO Coupling "External" mode, disables all parame-
ters in the "Frequency/Phase/Ref Osc" menu.
SCPI command:
Out State - LO Coupling
Switches on/off the LO output.
"On"
The internal local oscillator signal is also available on the LO OUT con-
nector (in order to couple two instruments).
"Off"
The LO OUT signal is switched off.
SCPI command:
5.4.6 NRP-Z Power Viewer
The R&S signal generator supports R&S NRP-Z power sensors to measure the power
of the output signal in the RF path or any freely selectable source. The generator performs
up to four power measurements, with the sensors either directly connected, or via the
R&S NRP-Z5 USB Sensor Hub.
RF Signal and Analog Modulations - RF Block