Remote Control Commands
R&S
®
SMBV100A
628
Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe
<State>
Switches on/off the LO output.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
FREQ:LOSC:OUTP:STAT ON
switches on the LO output state.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency[:CW|:FIXed]
<Fixed>
The command sets the frequency of the RF output signal for CW mode
(
SOURce:FREQuency:MODE CW
). In Sweep mode (
SOURce:FREQuency:MODE
SWEep
), this value is linked to the current sweep frequency.
In addition to a numerical value, it is also possible to specify
UP
and
DOWN
. The frequency
is then increased or decreased by the value which is set under
[
SOURce<[1]|2>:]FREQuency:STEP
.
As with the "FREQ" value entered in the display, the
OFFSet
value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet
is set to 0. The value range for other
OFFset
values can be calculated using
the following formula:
300 kHz +
OFFSet
... RF
max
+
OFFSet
Correlation
:
FREQ
for
FREQ:MODE SWE
is linked to sweep frequency.
Parameters:
<Fixed>
float
Range:
300kHz to RFmax
Increment: 0.01Hz
*RST:
1GHz
Default unit: Hz
Example:
FREQ 500kHz
sets the frequency of RF output signal A to 500 kHz.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency[:CW|:FIXed]:RCL
<Rcl>
The command determines whether the current frequency setting is retained or whether
the stored frequency setting is adopted when an instrument configuration is loaded.
*RST does not affect this setting.
SOURce Subsystem