Remote Control Commands
R&S
®
SGU100A
148
User Manual 1176.7726.02 ─ 09
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the
TRIGger
subsystem.
Usage:
Event
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
11.3
General Commands
:REMote:OPMode
<OpMode>
Sets the remote operation mode of the instrument.
Parameters:
<OpMode>
STDalone | EXTension
STDalone
The instrument acts standalone.
General Commands