Remote Control Commands
R&S
®
Scope Rider RTH
314
User Manual 1326.1578.02 ─ 12
●
Positive going pulses. The width is defined from the rising to the falling edge.
●
Negative going pulses. The width is defined from the falling to the rising edge.
●
Both positive and negative going pulses
Parameters:
<Polarity>
POSitive | NEGative | EITHer
*RST:
POSitive
TRIGger:WIDTh:RANGe
<Condition>
Defines how the measured pulse width is compared with the given limit(s).
Parameters:
<Condition>
LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers on pulses shorter or longer than a width set using
EQUal | NEQual
Triggers on pulses equal or not equal a given width that is set
using
. In addition, a tolerance can be
set around the specified width using
WITHin | OUTSide
Triggers on pulses inside or outside a given range. The range is
set using
and
*RST:
LONGer
TRIGger:WIDTh:WIDTh
<Duration>
Sets the width for comparison ranges EQUal, UNEQual, SHORter, and LONGer.
See
Parameters:
<Duration>
Range:
800E-12 to 10000
Increment: 100E-9
*RST:
5E-9
Default unit: s
TRIGger:WIDTh:DELTa
<Tolerance>
Sets a range
Δ
t to the specified width, which is defined using
.
Parameters:
<Tolerance>
Range:
0 to 5000
Increment: 500E-12
*RST:
0
Default unit: s
Waveform Setup