Operating the Instrument
R&S
®
RTO6
58
Getting Started 1801.6670.02 ─ 02
Cursor
The cursor icon shows the last selected cursor type. A short tap on the
icon activates the selected cursor.
To use another cursor type, select it in the toolbar assist, and adjust the settings.
Tap the diagram where you want to set the cursors, or draw a rectangle in the dia-
gram to position the cursor lines. The resulting cursor lines measure the selected
waveform, and the results are shown. You can adjust the cursor source, type and
position in the result box. Move the cursor lines by dragging them in the diagram,
or by turning the navigation knob. Pressing the knob switches the parameter to be
changed.
Mask Test
Starts the on-screen mask definition and the testing against the defined
mask.
Tap the icon and then tap the points that build the mask. To finish the shape defi-
nition, tap "Finish segment" in the toolbar assist. To create a rectangular mask,
draw a rectangle on the screen. You can add more segments to the test, and add
another test. You can also move the mask segments on the screen.
To configure the mask test settings, tap the icon in the "Mask" result box.
Histogram
The histogram icon on the toolbar shows the last selected histogram type.
A short tap on the icon activates the selected histogram.
To use another histogram type, select it in the toolbar assist, and adjust the
source.
Tap the icon and then drag a rectangle on the diagram to mark the histogram
area. The histogram for the selected waveform appears.
To change settings, touch and hold the histogram area, and the histogram toolbar
assist opens.
Measure
Starts an automatic measurement.
You can run up to 8 automatic measurement groups in parallel. The "Automatic
measurment" icon starts the measurement groups one after the other.
A measurement group consists of minimum one measurement, and can consist of
many measurements of the same category. Tap the icon, and select the category
and the measurements in the toolbar assist. Tap the diagram with the waveform
to be measured. To define a measurement gate, draw a rectangle on the screen.
Toolbar